Calibration

Also called: calibrate · two-point calibration · offset and gain · zeroing

Definition

Calibration compares an instrument's readings with known references and builds a correction, often an offset and a gain, so that future readings are accurate.

Picture it

Weigh nothing, then a known 1 kg mass. The empty reading is the offset; the change per kilogram is the gain. Any reading can then be corrected with a straight line. Re-check calibration over time, because sensors drift.

The maths

true = (raw − raw_zero) × (ref_value / (raw_ref − raw_zero))
SI units used in this entry
QuantityUnitSymbol
ratiodimensionless1

Worked example

A load cell reads raw 812 with nothing on it and 1612 with a 2.000 kg reference. What mass corresponds to a raw reading of 1212?

  1. Gain = 2.000 / (1612 − 812) = 0.0025 kg per count
  2. Mass = (1212 − 812) × 0.0025 = 1 kg

Answer1 kg (1.000 kg).

✓ The number in this answer is recomputed from scratch by an automated test.

Common misconceptions

  • Misconception: Calibrating at one point fixes the offset but not the gain.
  • Misconception: A calibration is only valid inside the range of the reference values used.